In search of the true dosage: Statistics applied to the analysis of the cocaine
A la recherche du juste dosage : statistique appliquée à l'analyse de la cocaïne
Résumé
A method of separation by gas chromatography with a flame ionization detector was developed for quantifying cocaine in powders seized by the police. The method was validated by studying parameters of calibration, trueness, precision based on trueness error (or systematic bias) and random error. Total error, which is the combination of these errors, is used to confirm the method adequacy with the objectives fixed by the analyst. Accuracy profile is an efficient decision tool to do it. Results obtained with weighted regression model allow concluding that the method fits quantitation of heroin and cocaine in powders on 2 to 100% concentration (w/w) domain with 10% limits of acceptation and a risk of 5%.
Une méthode de séparation par chromatographie gazeuse couplée à un détecteur à ionisation de flamme (CPG-FID) a été mise au point pour le dosage de la cocaïne dans des poudres issues de saisies policières. La méthode a été validée en étudiant les paramètres d’étalonnage, de justesse, de fidélité en se basant sur l’estimation de l’erreur de justesse (ou biais systématique) et de l’erreur aléatoire dont la combinaison (l’erreur totale) permet de vérifier si la méthode remplit les objectifs fixés par l’analyste. Pour cela, le profil d’exactitude s’avère être un outil de décision efficace. Les résultats obtenus avec un modèle de régression pondérée permettent de conclure que la méthode peut être appliquée au dosage de la cocaïne dans les poudres sur un domaine de concentrations (m/m) allant de 2 à 100 % pour des limites d’acceptation de 10 % avec un risque de 5 %.