Impedimetric aptasensor for thrombin detection - Université Claude Bernard Lyon 1 Accéder directement au contenu
Communication Dans Un Congrès Année : 2011

Impedimetric aptasensor for thrombin detection

Résumé

The goal of this work is to develop an impedimetric aptasensor sensitive, selective and stable for detection of thrombin. Thiol modified aptamer sequence being grafted on gold electrode, the faradic electrochemical impedance measurements allow the direct detection of thrombin in the range of 3.1-200 ngmL(-1) with a detection limit of 3.1 ngmL(-1) (80 fM). This study demonstrates that these aptasensors exhibit high sensitivity, high selectivity (compared to a model protein, BSA), stability and reproducibility.

Dates et versions

hal-00816512 , version 1 (22-04-2013)

Identifiants

Citer

N. Meini, Carole Farre, Carole Chaix, K. Kherrat, Sergei V. Dzyadevych, et al.. Impedimetric aptasensor for thrombin detection. 25th Eurosensors Conference, Sep 2011, Athens, Greece. ⟨10.1016/j.proeng.2011.12.361⟩. ⟨hal-00816512⟩